Description
BIM-6809 series NIR spectrometer is an upgraded version of BIM-6805 series. It adopts Hamamatsu uncooled 256 pixel InGaAs detector, which not only doubles the number of pixels, but also increases the pixel size, and the resolution and sensitivity have been greatly improved. It also adopts a replaceable slit design, a spectrometer can be configured with various slits of different widths to meet the spectrum measurement needs of different resolutions and sensitivities, making it possible for one machine to be used for multiple purposes. With its small size and flexible configuration, it is very suitable for industrial integration and scientific research analysis fields such as food sorting, biopharmaceuticals and life sciences.
Model List
Model | Wavelength Range | Sub Model | Resolution | Slit | Grating | Filter | Lens |
BIM-6809-01 | 920nm-1700nm | S03L00F00G37 | ~5nm | 25μm | 300g/mm,1200nm | -- | -- |
BIM-6809-02 | 1530nm-1705nm | S03L00F00G39 | ~1nm | 25μm | 1000g/mm,1310nm | -- | -- |
BIM-6809-03 | 950nm-1170nm | S03L00F00G39 | ~1nm | 25μm | 1000g/mm,1310nm | -- | -- |
*Other wavelength ranges and resolutions can be customized.
*Cylinder lenses can be added to some models to increase the sensitivity of the spectrometer.
Features
Adopt Hamamatsu uncooled 256-pixel InGaAs detector with extremely low power consumption;
Compact optical path and structure design, easy for system integration;
Compact optical path and structure design, small and light, easy for system integration and outdoor measurement;
Selectable wavelength range and resolution, reasonable configuration of spectral range and resolution;
Professional testing software, providing various professional measurement modules such as absorbance, transmittance, concentration and reflectance.
Application
920nm-1700nm light source measurement, including wavelength stability, intensity stability, and half-height width of light source
Food and agriculture: freshness and sugar testing of fruits and vegetables; fat and protein content testing of meat; quality analysis of grains and seeds, etc.
Composition analysis of oil
Life sciences and pharmaceuticals, pharmaceutical ingredient analysis and process monitoring
Component analysis, material screening and quality control in the production process of plastic products
Typical Data
Model | BIM-6809 |
Detector | InGaAs 256 pixel uncooled detector |
Pixel size 50µm×500µm | |
Dimension | 94mm×60mm×36.5mm |
Wavelength Range | 920nm-1700nm (pre-configured) |
Optical Resolution | ~5nm (pre-configured 25µm slit) |
A/D | 16bit |
Integration Time | 0.1ms-65s |
Trigger Mode | 4 types (normal trigger, software trigger, hardware trigger, synchronous trigger) |
Power Consumption | 5VDC,300mA |
Fiber Connector | SMA905, replaceable slit design |
Operating Temperature | 5℃-35℃ (recommended temperature 25℃) |
Communication Interface | USB2.0, RS232 |
Operating System | Win XP, Win 7, Win 8, Win10 |
Power Supply | Micro USB |
Typical Data
Xenon lamp spectrum test comparison [The optical resolution of BIM-6809 is better than that of BIM-6805]
Tungsten lamp spectrum test comparison [The sensitivity of BIM-6809 is better than that of BIM-6805]
Packing List
One spectrometer, one USB data cable, one U disk (including software and user manual), one calibration report, and one suitcase.
Software Download