$ 680.00+
SIM-6205 series Flash Xenon Light Source, 185-2000nm, Arc Length 3mm, 2W/5W
Power
  • Details


Overview

The Flash Pulse Xenon lamp is a new type of light source that integrates a xenon lamp bulb, a trigger, and a high-voltage power supply. It is widely used in spectral

measurement applications, because it does not require preheating and has low heat generation, it is more integrated into industrial online inspection systems.


Features

  • High radiation intensity and high flash instantaneous power are beneficial to improve the signal-to-noise ratio of the instrument

  • Wide spectral range, covering UV, visible and infrared, range 185-2000nm

  • Long service life, more than 100 million flashes

  • Good stability, CV value of 1000 flashes ≤ 2%

  • Small size, low heat generation, good for light machine stability

  • No need to warm up, start immediately, saving test time

  • DB-9 connector, plug and play


Application

  • UV, visible and infrared spectroscopic analysis

  • Fluorescence analysis

  • Medical treatment: Enzyme immunoassay, urine formed imager, small POCT biochemical instrument

  • Industrial automation strobe light source

  • Analysis of air pollution gas NO and SO2

  • Laboratory and online water pollution analysis


Parameter

Model

SIM-6205-0230

SIM-6205-0530

Power

2W

5W

Pluse Energy

8.8mJ-13.8mJ optional

22.4mJ - 35.0mJ optional

Arc Length

3.0 ± 0.3 mm

Transparent panel material

DM308 glass, UV transparent glass

Optical radiation wavelength range

185 - 2000 nm

Light output stability

Max 1.5% CV; Max4.2% PV -1000 flash

Life (number of flashes)

More than 100 million times

Input voltage

DC -Min 11.0 - Max 28.0 V

Input Current

11 V:Max 1.4 A

Trigger voltage

5V square wave pulse; high level pulse width ≥10 μS

Ambient temperature

0 - +40°C

Storage temperature

-40 - +90°C

Use and storage humidity

Max 95% Rh

Cooling method

Natural cooling

Weight

250g

Size

100 x 45 x 36mm


Typical spectrum

PIC01_光谱分布.jpg

Use Hitachi UH5300 spectrophotometer to measure the 200-1100nm spectral range


Light radiation stability

PIC02_光辐射稳定性.jpg

Use Hitachi UH5300 spectrophotometer to measure 600s


Light intensity distribution curve

PIC_光强分布.jpg


Electrical input DB-9 wiring diagram

PIC_电气输入DB-9接线示意图.png


Working sequence diagram

PIC_触发时序图.png


Fixed hole size drawing

PIC_固定孔位图.jpg